Cartridge & Rotor
Operating point
Main solute A
Tracer B (optional)
Main solute A across the flux range — same mean ΔP, solution and cartridge, same B
| Jv (LMH) | 𝒢A | kA (µm/s) | Jv/kA | ΓA | cm,A (g/L) | Rint,A | Robs,A |
|---|
Tracer B across the flux range — same conditions, same BB
| Jv (LMH) | kB (µm/s) | Jv/kB | ΓB | cm,B (mg/L) | Rint,B | Robs,B |
|---|
Solution properties used
Want to see how this compares with a stirred cell (Sterlitech HP4750) or a spiral-wound element at the same ΔP? Open the device comparison.
Model: Geraldes V., Fan H., Yang M., Rodrigues M., Elimelech M. — "Predicting concentration polarization in Ortho-Centrifugal Membrane Filters for NF/RO membrane screening" (2026): kc/U* = 0.28 𝒢1/4 Sc−1/4, U* = (amD)1/3, 𝒢A = RA βA cA0 Jv/UA* (Eq. 35–36); a dilute tracer B follows kB/UB* = 0.28 𝒢A1/4 ScB−1/4 (Eq. 47) — the convective field is set by A alone, B only contributes its own diffusivity. Electrolyte properties are estimated from ionic data: D from limiting ionic conductivities (Nernst–Hartley, with a Davies thermodynamic factor and a viscosity correction), viscosity from Jones–Dole ionic B-coefficients, β = (1/ρ) dρ/dc from conventional partial molar ionic volumes (Millero 1972); a trace ion in a supporting electrolyte uses its own ionic diffusivity. Neutral solutes: literature D° and partial molar volumes. All data at 25 °C (CRC Handbook; Jenkins & Marcus 1995), scaled to the set temperature. The framework was derived for neutral solutes; for electrolytes it is applied with the salt (Nernst–Hartley) diffusivity. For NaCl, MgSO₄ and lactose the device comparison uses measured property fits instead. UF model (macromolecules): the same law with its prefactor re-extracted from the DNS with the finite-permeation closure of Geraldes & Afonso (2006), Γ = Robs φ (1 + 0.26 φ1.4)1.7, replacing film theory (valid only for φ ≲ 1.5), and with ν and D evaluated at the film concentration ½(cm + c0); validated on twelve reconstructed PEG 35 kDa ultrafiltration experiments (Pereira, MSc thesis, IST 2026; Sc ≈ 3×10⁴, Γ 2–66, φ 1.4–7, within ±20 % in Γ). Properties of PEG 35 kDa from measured fits at 20 °C (Pires et al.). Osmotic pressure π = φ ν m R T ρw with the osmotic coefficient φ from the Pitzer ion-interaction model (β⁽⁰⁾, β⁽¹⁾, β⁽²⁾, Cφ at 25 °C from Kim & Frederick, J. Chem. Eng. Data 33 (1988) 177, for ~140 salts; class-average parameters, flagged, otherwise) and φ = 1 + b·m for neutral solutes; it is used only to bound the input (πfeed ≤ 50 bar) and to check that the set flux is compatible with ΔP. Pressure: ΔP = ½ρω²(rm² − rs²) with the feed density ρ and the free-surface radius rs(V) from the cartridge calibration; since rs increases as the feed volume drops, the rotor speed is chosen for the arithmetic mean of ΔP at the initial volume V0 and at the final volume V0 − Vp (within ~0.3 bar of the time-averaged pressure for Vp up to ~half of V0).